info@universallab.org
WhatsApp: +41762172997
logo-header
​
logo
Material Testing
Mechanical Testing
Tensile TestFibre Tensile TestHardness TestImpact TestFatigue TestOthers
Thermal Testing
Thermomechanical AnalysisDSCTGATGA (TG-DSC, TG-DTA)
Surface, Coating and Corrosion
Adhesion TestScratch TestCorrosion Cycle Test
Physical Properties
Functional propertiesPorosityDensity
Elemental & Chemical Composition
Compositional analysisElemental analysis
Microscopic Characterization
Particle analysisMicroscopy
Product Testing
Food & Agriculture Testing
Consumer Goods Testing
Electronics & Electrical Testing
Medical & Pharmaceutical Testing
Automotive & Transportation Testing
Packaging & Materials Testing
Chemical & Hazardous Substances Testing
Semiconductor Testing
Battery Testing
Petrochemical Testing
Environmental Test
Water Testing
Drinking Water Analysis
Soil & Sediment Testing
Soil Chemistry (pH, Nutrients)
Air Testing
Particulate Matter (PM2.5, PM10)
Noise & Acoustic Testing
Industrial/Occupational Noise
Waste & Sludge Testing
Hazardous Waste Classification
Instrumental Test
Spectroscopy
Emission
XRFXPSICP-OESGD-OES
Absorption
FTIRUV/VIS/NIRAAS
Mass Spectrometry
SIMSICP-MS
Other
RamanNMR
Microscopy
Electron
SEMTEMEBSDEMPA
Other
AFMFIB
Mechanical
Volumetric
Tensile TestFibre Tensile TestImpact TestFatigue Test
Surface
HardnessNanoindentationScratch
Others
Thermal Analysis
TGADSC
Physical Analysis
BETLaser diffractionDensityContact AngleZeta potential
Chromatography
GCHPLCIC
Services
  • Applications
  • Techniques
  • Standards
  • Industries
Regulatory testing
  • PFAS
  • REACH
  • VOC
Environmental testing
  • Water
  • Gas
Data analysis
  • XRD
  • XPS
  • FTIR
  • Raman
  • NMR
  • CT
  • SIMS
  • TGA
Blog
  • Knowledge
  • News
  • Policy
  • About us
    Contact
    Published on
    Saturday, March 8, 2025

    Principles and Techniques of X-Ray Diffraction (XRD)

    Authors
    • avatar
      Name
      Universallab
      Twitter
      @universallab

    Basic Principles of XRD

    The basic principle of X-ray diffraction is based on Bragg's Law: nλ = 2d·sinθ, where:

    • n is the diffraction order (usually 1)
    • λ is the wavelength of the incident X-ray
    • d is the interplanar spacing
    • θ is the incident angle

    When X-rays are directed at a crystalline sample, atoms in the crystal scatter the X-rays. At specific angles, constructive interference occurs between scattered waves, producing strong diffraction peaks. By measuring the positions and intensities of these peaks, the crystal structure of the material can be determined.

    Bragg's Diffraction Condition

    When X-rays with an angle θ strike parallel crystal planes with spacing d, diffraction peaks occur when the path difference between X-rays reflected from adjacent planes equals an integer multiple (nλ) of the wavelength. This is the famous Bragg equation.

    Interpreting Diffraction Patterns

    Each peak in a diffraction pattern corresponds to diffraction from specific crystal planes. The peak position (2θ value) is related to the interplanar spacing, while the peak intensity is related to the types and arrangement of atoms on the plane. By comparing with standard diffraction pattern databases (such as JCPDS cards), the phase composition of the sample can be determined.

    XRD Instrumentation

    Modern XRD instruments primarily consist of the following components:

    1. X-ray source: Typically copper target (Cu Kα, λ=1.5418Å) or molybdenum target (Mo Kα, λ=0.7107Å)
    2. Collimation system: Controls the divergence of the X-ray beam
    3. Sample stage: Can rotate to change the incident angle of X-rays
    4. Detector: Receives diffracted X-rays and records intensity
    5. Data acquisition and processing system: Converts detected signals into analyzable diffraction patterns

    Depending on different analytical purposes, XRD instruments can be categorized as powder diffractometers, single crystal diffractometers, thin film diffractometers, and more.

    XRD Sample Preparation Techniques

    The quality of sample preparation directly affects the accuracy of XRD analysis. Here are some key sample preparation techniques:

    Powder Sample Preparation

    1. Particle size control: Samples should be ground to below 10μm to reduce preferred orientation and micro-absorption effects
    2. Uniform filling: Samples should be uniformly filled in the sample holder with a flat surface
    3. Avoid over-compaction: Excessive compaction leads to preferred orientation, affecting peak intensities
    4. Appropriate amount: Sample thickness should be sufficient to prevent X-rays from penetrating to the sample holder, but not excessively thick

    Thin Film Sample Preparation

    1. Substrate selection: Choose substrates with low background and no interfering peaks
    2. Surface cleanliness: Ensure clean substrate and film surfaces without contamination
    3. Thickness control: Film thickness should be moderate; too thin results in weak signals, too thick may produce stress

    Single Crystal Sample Preparation

    1. Crystal selection: Choose crystals with good transparency, no cracks, and regular shapes
    2. Size control: Crystal sizes are typically ideal between 0.1-0.4mm
    3. Careful mounting: Use nylon loops or glass fibers to mount crystals without damaging them

    Optimization of XRD Test Parameters

    Setting appropriate test parameters is crucial for obtaining high-quality diffraction data:

    Scan Rate Selection

    1. Phase identification: Fast scanning (5-10°/min) is sufficient
    2. Fine structure analysis: Slow scanning (0.5-2°/min) for higher resolution
    3. Trace phase detection: Very slow scanning (0.1-0.5°/min) with increased counting time per point

    Scan Range Setting

    1. Routine phase analysis: 10-80° (2θ) is usually sufficient
    2. Low-angle diffraction materials: Need to start scanning from lower angles (e.g., 3-5°)
    3. Fine structure analysis: May need to extend to high angles (>100°)

    Step Size and Counting Time

    1. Step size: 0.02° is typically sufficient for phase identification; fine structure analysis requires 0.01° or smaller
    2. Counting time: Depends on the sample's scattering ability; trace phases or weak diffraction require longer counting times

    XRD Data Analysis Techniques

    After obtaining diffraction data, effective analysis is key to extracting useful information:

    Phase Identification

    1. Background subtraction: Choose appropriate background subtraction methods, avoiding artificial introduction or elimination of features
    2. Peak position determination: Use suitable peak fitting methods to determine accurate peak positions
    3. Database searching: Combine d-values, relative intensities, and chemical information for database searching
    4. Qualitative analysis: Ensure that all major diffraction peaks have corresponding explanations

    Quantitative Analysis

    1. Direct comparison method: Suitable for simple binary systems
    2. Internal standard method: Add standard materials of known content as references
    3. Full pattern fitting method (e.g., Rietveld method): Suitable for complex multiphase systems
    4. Considerations: Preferred orientation, micro-absorption effects, etc., affect the accuracy of quantitative analysis

    Lattice Parameter Determination

    1. High-angle peak selection: High-angle peaks are more sensitive to lattice parameter changes
    2. Internal standard correction: Use internal standards to eliminate systematic errors
    3. Extrapolation methods: Methods such as the Nelson-Riley function can improve precision

    Advanced Applications of XRD

    Beyond basic phase analysis, XRD has many advanced applications:

    Crystallite Size and Microstrain Analysis

    Using the Scherrer formula or Williamson-Hall method to analyze peak width can estimate crystallite size and microstrain. Scherrer formula: D = Kλ/(β·cosθ), where K is the shape factor and β is the full width at half maximum of the peak.

    Texture Analysis

    Pole figure measurements can determine the orientation distribution of grains in a material, which is crucial for understanding anisotropic properties of materials.

    Stress Analysis

    Based on the principle that interplanar spacing changes with stress state, residual stress in materials can be calculated by measuring changes in diffraction angles of specific crystal planes.

    In-situ XRD Analysis

    By combining heating, cooling, tensile, or atmosphere control devices, researchers can study structural evolution processes of materials under different conditions.

    Common Problems and Solutions

    Various problems may be encountered in XRD experiments. Here are some common issues and solutions:

    High Background

    Causes: Amorphous components in samples, fluorescence effects, sample holder contamination Solutions: Use monochromators, reduce exposure area, clean sample holders, select appropriate filters

    Peak Position Shift

    Causes: Sample height errors, instrument calibration issues, sample displacement Solutions: Add internal standards, correctly adjust sample height, regularly calibrate instruments

    Abnormal Peak Intensities

    Causes: Preferred orientation, sample inhomogeneity, insufficient statistics Solutions: Rotate samples, repeat sample loading and average results, extend counting time, use ball milling for thorough grinding

    Severe Peak Overlap

    Causes: Complex multiphase systems, low symmetry structures Solutions: Use high-resolution mode, switch to X-ray sources with shorter wavelengths, use peak fitting software, consider synergistic use of other characterization methods

    Conclusion

    As one of the cornerstones of material characterization, X-ray diffraction technology continues to expand its application range. Mastering the basic principles and techniques of XRD is crucial for materials science researchers. Through proper sample preparation, parameter optimization, and data analysis, XRD can provide rich and accurate information for material structure research. With the development of advanced light sources such as synchrotron radiation, XRD technology will continue to play a key role in materials science.

    References

    1. Cullity, B. D., & Stock, S. R. (2014). Elements of X-ray diffraction (3rd ed.). Pearson Education.
    2. Pecharsky, V. K., & Zavalij, P. Y. (2009). Fundamentals of powder diffraction and structural characterization of materials (2nd ed.). Springer.
    3. He, B. B. (2018). Two-dimensional X-ray diffraction (2nd ed.). John Wiley & Sons.
    4. Dinnebier, R. E., & Billinge, S. J. (Eds.). (2008). Powder diffraction: Theory and practice. Royal Society of Chemistry.
    5. Waseda, Y., Matsubara, E., & Shinoda, K. (2011). X-ray diffraction crystallography: introduction, examples and solved problems. Springer.
    Discuss on LinkedIn

    Tags

    ← Back to the blog
    HomeCategoryCart
    logo-footer-shop1

    European material analytical and measuring service provider dedicated for academic and industrial research

    About Us
    CareersTerms & ConditionsPrivacy Policy
    Customer Services
    Help CenterBecome a partnerReturns & Refunds
    Contact Us

    Parkstrasse 1, 5234 Villigen, Switzerland

    Email: info@universallab.org

    Phone: +41 762172997

    VAT: CHE-265.266.739 MWST

    © 2026 By Universallab. All rights reserved.

    This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.